Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-265
limit1Pattern
The bit pattern that is sent when the measurement fails limit 1; range 1 to
15; default is 1
limit2Pattern
The bit pattern that is sent when the measurement fails limit 2; range 1 to
15; default is 2
limit3Pattern
The bit pattern that is sent when the measurement fails limit 3; range 1 to
15; default is 4
limit4Pattern
The bit pattern that is sent when the measurement fails limit 4; range 1 to
15; default is 8
allPattern
The bit pattern that is sent when all limits have passed; 1 to 15; default
is 15
bufferName
The name of the reading buffer, which may be a default buffer
(
defbuffer1
or
defbuffer2
) or a user-defined buffer; defaults to
defbuffer1
Details
This trigger model template allows you to grade components and place them into up to four bins,
based on the comparison to limits.
To set a limit as unused, set the high value for the limit to be less than the low limit.
All limit patterns and the pass pattern are sent on digital I/O lines 1 to 4, where 1 is the least
significant bit.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel
MENU
key and under Trigger, selecting
Configure
. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist()
command to view the trigger model blocks in a list format.
Also see
None