Section 2: General operation
Model 2461 Interactive SourceMeter® Instrument Reference Manual
2-30
2461-901-01 A/November 2015
Measure Config Lists menu
The
Config Lists
menu allows you to select an existing measure configuration list,
create a new list, load configuration settings to and from the instrument (system), and
view the settings of an index in a configuration list. For more information about using
configuration lists, see
(on page 3-30).
Setting
Description
Select List
Displays a list of available configuration lists from which you can
choose. If you selected
Source
on the main menu, a menu of the
source configuration lists presently available on the instrument is
displayed.
If you selected
Measure
on the main menu, a menu of the
measure configuration lists presently available on the instrument is
displayed.
New List
Creates a new, empty configuration list. To populate the list with
the present instrument settings, select
System to Index
.
Delete Index
Deletes a configuration list index from the selected configuration
list.
System to Index
Saves the present instrument settings to an index at the end of the
selected configuration list.
Index to System
Restores the instrument to the settings stored in the selected
configuration list index.
Index Details
Displays the details of the selected configuration index. Details
include settings such as function, value, delay, calculations, and
range. You can scroll the displayed list to view additional settings.