Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-310
2461-901-01 A/November 2015
Details
If one or more trigger events were detected since the last time
trigger.timer[
N
].wait()
or
trigger.timer[
N
].clear()
was called, this function returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed.
This is true regardless of the number of events detected.
Example
triggered = trigger.timer[3].wait(10)
print(triggered)
Waits up to 10 s for a trigger on timer 3.
If
false
is returned, no trigger was detected
during the 10 s timeout.
If
true
is returned, a trigger was detected.
Also see
(on page 8-301)
trigger.tsplinkin[N].clear()
This function clears the event detector for a LAN trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.tsplinkin[
N
].clear()
N
The trigger line (1 to 3) to clear
Details
The trigger event detector enters the detected state when an event is detected. When this command
is sent, the instrument does the following actions:
•
Clears the trigger event detector
•
Discards the history of the trigger line
•
Clears the
trigger.tsplinkin[
N
].overrun
attribute
Example
tsplink.line[2].mode =
tsplink.MODE_TRIGGER_OPEN_DRAIN
trigger.tsplinkin[2].clear()
Clears the trigger event on TSP-
Link line 2.
Also see
(on page 8-311)
(on page 8-319)