Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2461-901-01 A/November 2015
3-61
Using the front panel:
Press the front-panel
TRIGGER
key for two seconds and select
Abort Trigger Model
.
Using SCPI commands:
Send the command:
:ABORt
Using TSP commands:
Send the command:
trigger.model.abort()
Sweep programming examples
The following examples show programming examples of typical sweeps.
Linear sweep with a voltage source
The following examples perform a linear sweep that uses a voltage source. They perform the
following actions:
•
Reset the instrument to its defaults.
•
Set the source function to voltage.
•
Set the source range to 20 V.
•
Set the source limit for measurements to 0.02 A
•
Set the measure function to current.
•
Set the current range to automatic.
•
Set up a linear sweep that sweeps from 0 to 10 V in 21 steps with a source delay of 200 ms.
•
In TSP only, name the configuration list that is created for this sweep
RES
.
•
Start the sweeps.
•
Wait until all commands are complete and then query the source value and measurement
reading.
No buffer is defined, so the data is stored in
defbuffer1
(on page 3-2) for
more information about reading buffers.
Using SCPI commands
*RST
SOUR:FUNC VOLT
SOUR:VOLT:RANG 20
SOUR:VOLT:ILIM 0.02
SENS:FUNC "CURR"
SENS:CURR:RANG:AUTO ON
SOUR:SWE:VOLT:LIN 0, 10, 21, 200e-3
INIT
*WAI
TRAC:DATA? 1, 21, "defbuffer1", SOUR, READ