Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-225
:TRIGger:LOAD "Empty"
This command resets the trigger model.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:LOAD "Empty"
Details
When you load this predefined trigger model, any blocks that have been defined in the trigger model
are cleared so the trigger model has no blocks defined.
This command replaces the
TRIGger:LOAD:EMPTy
command, which is deprecated.
Example
TRIG:LOAD "Empty"
TRIG:BLOC:BUFF:CLEAR 1
TRIG:BLOC:MEAS 2
TRIG:BLOC:BRAN:COUN 3, 5, 2
TRIG:BLOC:DEL:CONS 4, 1
TRIG:BLOC:BRAN:COUN 5, 3, 2
Reset trigger model settings.
Clear
defbuffer1
at the beginning of execution
of the trigger model.
Loop and take 5 readings.
Delay 1 s.
Loop three more times back to block 2.
At the end of execution, 15 readings are stored
in
defbuffer1
.