Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-37
Example
reset()
testData = buffer.make(50)
smu.source.output = smu.ON
trigger.model.load("SimpleLoop", 2, 0, testData)
trigger.model.initiate()
waitcomplete()
printbuffer(1, 2, testData.statuses)
Create a reading buffer named
testData
,
configure the instrument to make two
measurements, and store the readings in
the buffer.
Turn on the source output.
Print the source status for the readings in
testData
.
Output:
64, 64
Indicating that the status is
buffer.STAT_LIMIT1_LOW
.
Also see
(on page 8-14)
(on page 8-11)
(on page 8-19)
(on page 8-31)
(on page 8-92)
(on page 8-93)
(on page 3-2)
(on page 3-23)
bufferVar.times
This attribute contains the time when the instrument made the reading.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Restore configuration
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
readingTime
=
bufferVar
.times[
N
]
readingTime
The time of the reading in hours, minutes, and seconds
bufferVar
The name of the reading buffer, which may be a default buffer (
defbuffer1
or
defbuffer2
) or a user-defined buffer
N
The reading number
N
; can be any value from 1 to the number of readings in the
buffer; use the
bufferVar.
n
command to determine the number of readings in the
buffer