Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-297
Example
trigger.model.setblock(2, trigger.BLOCK_SOURCE_OUTPUT, smu.ON)
Set trigger model to turn the source on when it reaches block 2.
Also see
(on page 3-110)
trigger.model.setblock() —
trigger.BLOCK_SOURCE_PULSE_OUTPUT
This function defines a pulse trigger block that turns the pulse source on or off.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(
blockNumber
, trigger.BLOCK_SOURCE_PULSE_OUTPUT,
state
)
blockNumber
The sequence of the block in the trigger model
state
Turn the source pulse off:
smu.OFF
Turn the source pulse on:
smu.ON
Details
The source pulse output block determines if the pulse output source is turned on or off when the
trigger model reaches this block.
The source output must be on before the source pulse output is turned on. If the source output is not
on when the trigger model reaches the source pulse output block in the trigger model, a pulse is not
generated. To avoid this, either enable the source output before initiating the trigger model, or make
sure there is a source output block in the trigger model before the source pulse output block.
This block does not determine the settings of the pulse output source (such as the pulse level and
pulse limit). The pulse source settings are determined by either of the following:
•
The present settings of the instrument
•
The recall of a source configuration list containing the settings for your application
When you list trigger blocks, this block is listed as
PULSE_OUTPUT
.
Example 1
trigger.model.setblock(4, trigger.BLOCK_SOURCE_OUTPUT, smu.ON)
trigger.model.setblock(5, trigger.BLOCK_SOURCE_PULSE_OUTPUT, smu.ON)
Set trigger model to turn the source on when it reaches block 4 and source pulse on when it reaches block 5.