![background image](http://html2.mh-extra.com/html/keithley/2461/2461_reference-manual_39936521103.webp)
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Appendix D: Convert Model 2400 to Model 2461 code
2461-901-01 A/November 2015
D-31
Model 2400 command
:TRIGger[:SEQuence[1]][:TCONfigure]:DIRection
:TRIGger[:SEQuence[1]][:TCONfigure]:DIRection?
Model 2461 command
Not available
Notes
Similar functionality available with
(on page 6-209).
Model 2400 command
:TRIGger[:SEQuence[1]][:TCONfigure]:ILINe
:TRIGger[:SEQuence[1]][:TCONfigure]:ILINe?
Model 2461 command
Not available
Notes
For similar functionality, use a digital I/O line with the trigger model.
Generate a notify event in the trigger model that feeds the stimulus setting to a
digital I/O line to pulse as needed in the trigger model.
See:
•
(on page 6-206)
•
:TRIGger:DIGital<n>:OUT:STIMulus
(on page 6-214)
Model 2400 command
:TRIGger[:SEQuence[1]][:TCONfigure]:OLINe
:TRIGger[:SEQuence[1]][:TCONfigure]:OLINe?
Model 2461 command
Not available
Notes
For similar functionality, use a digital I/O line with the trigger model.
Generate a notify event in the trigger model that feeds the stimulus setting to a
digital I/O line to pulse as needed in the trigger model.
See:
•
(on page 6-206)
•
:TRIGger:DIGital<n>:OUT:STIMulus
(on page 6-214)
Model 2400 command
:TRIGger[:SEQuence[1]][:TCONfigure]:OUTPut
:TRIGger[:SEQuence[1]][:TCONfigure]:OUTPut?
Model 2461 command
Not available
Notes
Similar functionality available with
to a digital I/O line.
Model 2400 command
:TRIGger:SEQuence2:SOURce
:TRIGger:SEQuence2:SOURce?
Model 2461 command
Not available
Notes
Model 2400 command
:TRIGger:SEQuence2:TOUT
:TRIGger:SEQuence2:TOUT?
Model 2461 command
Not available
Notes