![background image](http://html2.mh-extra.com/html/keithley/2461/2461_reference-manual_39936521092.webp)
Appendix D: Convert Model 2400 to Model 2461 code
Model 2461 Interactive SourceMeter® Instrument Reference Manual
D-20
2461-901-01 A/November 2015
Model 2400 command
:SOURce[1]:DELay
:SOURce[1]:DELay?
Model 2461 command
(on page 6-89), where
<function>
is
CURRent
or
VOLTage
.
:SOURce[1]:<function>:DELay?
Notes
When a delay is set, source auto delay is turned off.
Model 2400 command
:SOURce[1]:DELay:AUTO
:SOURce[1]:DELay:AUTO?
Model 2461 command
:SOURce[1]:<function>:DELay:AUTO
(on page 6-90)
:SOURce[1]:DELay:AUTO?
Notes
Model 2400 command
:SOURce[1]:FUNCtion[:MODE]
:SOURce[1]:FUNCtion[:MODE]?
Model 2461 command
(on page 6-95)
:SOURce[1]:FUNCtion[:MODE]?
Notes
Model 2400 command
:SOURce[1]:LIST:CURRent
:SOURce[1]:LIST:CURRent?
Model 2461 command
:SOURce[1]:SWEep:<function>:LIST
(on page 6-128)
Notes
In the Model 2461, this sets up a list of customer values for a sweep.
Model 2400 command
:SOURce[1]:LIST:CURRent:APPend
Model 2461 command
:SOURce[1]:SWEep:<function>:LIST
(on page 6-128)
Notes
In the Model 2461, this setting is set as part of the configuration list that is
created by the sweep command.
Model 2400 command
:SOURce[1]:LIST:CURRent:POINts?
Model 2461 command
:SOURce[1]:SWEep:<function>:LIST
(on page 6-128)
Notes
In the Model 2461, this setting is set as part of the configuration list that is
created by the sweep command.
Model 2400 command
:SOURce[1]:LIST:CURRent:STARt
:SOURce[1]:LIST:CURRent:STARt?
Model 2461 command
:SOURce[1]:SWEep:<function>:LIST
(on page 6-128)
Notes
In the Model 2461, this setting is set as part of the configuration list that is
created by the sweep command.
Model 2400 command
:SOURce[1]:LIST:VOLTage
:SOURce[1]:LIST:VOLTage?
Model 2461 command
:SOURce[1]:SWEep:<function>:LIST
(on page 6-128)
Notes
In the Model 2461, this sets up a list of customer values for a sweep.
Model 2400 command
:SOURce[1]:LIST:VOLTage:APPend
Model 2461 command
:SOURce[1]:SWEep:<function>:LIST
(on page 6-128)
Notes
In the Model 2461, this setting is set as part of the configuration list that is
created by the sweep command.