Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-84
2461-901-01 A/November 2015
localnode.gettime()
This function retrieves the instrument date and time.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
localnode.gettime()
Details
The time is returned in UTC time. UTC time is specified as the number of seconds since Jan 1, 1970,
UTC. You can use UTC time from a local time specification, or you can use UTC time from another
source (for example, your computer).
Example
print(os.date('%c', gettime()))
Example output:
Wed Mar 31 14:25:31 2010
Also see
(on page 8-87)
localnode.linefreq
This attribute contains the power line frequency setting that is used for NPLC calculations.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Power cycle
Not applicable
Not applicable
Usage
frequency
= localnode.linefreq
frequency
The detected line frequency: 50 or 60
Details
The instrument automatically detects the power line frequency (either 50 Hz or 60 Hz) when the
instrument is powered on.
Example
frequency = localnode.linefreq
print(frequency)
Reads line frequency setting.
Also see
None