Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-21
bufferVar.endindex
This attribute indicates the last index in a reading buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Restore configuration
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
bufferVar
.endindex =
endIndex
endIndex
Ending index of the buffer
bufferVar
The name of the reading buffer, which may be a default buffer (
defbuffer1
or
defbuffer2
) or a user-defined buffer
Details
Use this attribute to find the ending index in a reading buffer.
Example
test1 = buffer.make(100)
smu.measure.count = 6
smu.measure.read(test1)
print(test1.startindex, test1.endindex, test1.capacity)
smu.measure.read(test1)
print(test1.startindex, test1.endindex)
Create a buffer named
test1
with a capacity of 100 readings.
Set the measure count to 6.
Make measurements and store them in buffer
test1
.
Get the start index, end index, and capacity of
test1
.
Output:
1, 6, 100
Make six more measurements and store them in buffer
test1
.
Get the start index and end index of
test1
.
Output:
1, 12
Also see
(on page 8-35)
(on page 8-14)
(on page 3-2)
(on page 3-23)