Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-234
2461-901-01 A/November 2015
Details
When this command returns zero (0), no bits are set. You can send 0 to clear all bits.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register.
Bit
Decimal
value
Constant
When set, indicates the following has occurred:
0
1
status.standard.OPC
All pending selected instrument operations are
complete and the instrument is ready to accept new
commands. The bit is set in response to an
(on page B-6) command or TSP
(on page 8-
92) function.
1
2
Not used
Not used.
2
4
status.standard.QYE
Attempt to read data from an empty Output Queue.
3
8
Not used
Not used.
4
16
Not used
Not used.
5
32
Not used
Not used.
6
64
Not used
Not used.
7
128
status.standard.PON
The instrument has been turned off and turned back
on since the last time this register was read.
Command errors include:
•
IEEE Std 488.2 syntax error:
The instrument received a message that does not follow the defined
syntax of the IEEE Std 488.2 standard.
•
Semantic error:
The instrument received a command that was misspelled or received an optional IEEE
Std 488.2 command that is not implemented in the instrument.
•
GET error:
The instrument received a Group Execute Trigger (GET) inside a program message.
Example
print(status.standard.event)
May return the value
129
, showing that the Standard Event
Status Register contains binary
10000001
Also see
(on page C-3)
(on page C-16)
timer.cleartime()
This function resets the timer to zero (0) seconds.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
timer.cleartime()