Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-184
2461-901-01 A/November 2015
smu.source.func
This attribute contains the source function, which can be voltage or current.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Configuration script
Source configuration list
smu.FUNC_DC_VOLTAGE
Usage
sFunction
= smu.source.func
smu.source.func =
sFunction
sFunction
The source function; set to one of the following values:
•
Current source:
smu.FUNC_DC_CURRENT
•
Voltage source:
smu.FUNC_DC_VOLTAGE
Details
When you set this command, it configures the instrument as either a voltage source or a current
source.
When you read this command, it returns the output setting of the source.
Example
smu.source.func = smu.FUNC_DC_CURRENT
Sets the source function of the
instrument to be a current source.
Also see
(on page 8-185)
(on page 8-188)
smu.source.highc
This attribute enables or disables high-capacitance mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Function change
Configuration script
Source configuration list
smu.OFF
Usage
state
= smu.source.highc
smu.source.highc =
state
state
Turn high-capacitance mode off:
smu.OFF
Turn high-capacitance mode on:
smu.ON