Section 3: Functions and features
Model 2461 Interactive SourceMeter® Instrument Reference Manual
3-52
2461-901-01 A/November 2015
To run autocalibration from the front panel:
1. Press the
MENU
key.
2. Under System, select
Calibration
.
3. Select
Start ACAL
. A prompt is displayed.
4. Select
Yes
. The Start ACAL button turns gray and a spinning activity indicator is displayed while
the calibration runs.
After the autocalibration is complete, the information on the Calibration screen is updated.
To run autocalibration using SCPI commands:
Send:
ACAL:RUN
To run autocalibration using TSP commands:
Send:
acal.run()
Once autocalibration has started, you cannot stop it.
Reviewing calibration information
The Calibration screen displays information about the last autocalibration and factory calibrations that
were run and the present status. For detail on this screen, refer to
2-45).
For autocalibration, you can also access this information from the commands in the SCPI
(on page 6-16) or the TSP
acal.*
commands.