Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2461-901-01 A/November 2015
3-63
Voltage sweep based on a configuration list
The following TSP example shows a voltage sweep that is based on a configuration list. It performs
the following actions:
•
Reset the instrument to its defaults
•
Create a source configuration list called
CurrListSweep
.
•
Set the source function to current.
•
Set the source current range to 100 mA.
•
Set the source current level to 1 mA.
•
Save the source settings to
CurrListSweep
.
•
Set the source current level to 10 mA.
•
Save the source settings to
CurrListSweep
.
•
Set the source current level to 5 mA
•
Save the source settings to
CurrListSweep
.
•
Set the source current level to 7 mA.
•
Save the source settings to
CurrListSweep
.
•
Set the source current level to 11 mA.
•
Save the source settings to
CurrListSweep
.
•
Set the source current level to 9 mA.
•
Save the source settings to
CurrListSweep
.
•
Set up a list sweep that uses the entries from the
CurrListSweep
configuration list and starts at
index 1 of the list.
•
Start the sweep.
No buffer is defined, so the data is stored in
defbuffer1
(on page 3-2) for
more information on reading buffers.