Section 3: Functions and features
Model 2461 Interactive SourceMeter® Instrument Reference Manual
3-48
2461-901-01 A/November 2015
Using SCPI commands:
:SOURce:CONFiguration:LIST:RECall "MySourceList", 2
For a measure configuration list, replace
:SOURce
with
:SENSe
and specify an existing measure
configuration list name and index.
Using TSP commands:
smu.source.configlist.recall("MySourceList", 2)
For a measure configuration list, replace
source
with
measure
and specify an existing configuration
list name and index.
The following figure shows the active state of the instrument after you recall configuration index 2.
Notice that the NPLC setting is 2.0. This is because when you recall a source configuration list, the
settings that are restored are the source settings that were active at the time the point was stored.
Since NPLC is not a source setting, recalling a source configuration list has no effect on the value of
the NPLC setting.
Figure 110: Recall configuration point 2
Viewing configuration list contents
You can ask the instrument to display or print the contents of a specific configuration index by
sending a query. The contents returned include all of the active settings that the instrument saved
when you stored the configuration index.