Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-262
2461-901-01 A/November 2015
Details
When you load this predefined trigger model, you can specify amount of time to make a
measurement and the length of the delay before the measurement.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel
MENU
key and under Trigger, selecting
Configure
. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist()
command to view the trigger model blocks in a list format.
Example
reset()
--set up measure function
smu.measure.func = smu.FUNC_DC_CURRENT
--set up source function
smu.source.func = smu.FUNC_DC_VOLTAGE
smu.source.level = 5
--turn on output and initiate readings
trigger.model.load("DurationLoop", 10, 0.01)
trigger.model.initiate()
Reset the instrument. Set the instrument to source voltage at 5 V. Set to measure current.
Load the duration loop trigger model to take measurements for 10 s with a 10 ms delay before each
measurement.
Start the trigger model.
Also see
None