Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-236
2461-901-01 A/November 2015
:TRIGger:TIMer<n>:COUNt
This command sets the number of events to generate each time the timer generates a trigger event or is enabled
as a timer or alarm.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
1
Usage
:TRIGger:TIMer<n>:COUNt <count>
:TRIGger:TIMer<n>:COUNt?
<n>
Trigger timer number (1 to 4)
<count>
The number of times to repeat the trigger (0 to 1,048,575)
Details
If
count
is set to a number greater than 1, the timer automatically starts the next trigger timer delay
at the expiration of the previous delay.
Set
count
to zero (0) to cause the timer to generate trigger events indefinitely.
If you use the trigger timer with a trigger model, make sure the count value is the same or more than
any count values expected in the trigger model.
Example 1
TRIG:TIM2:COUN 4
Set the number of events to generate for trigger