Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-210
2461-901-01 A/November 2015
The following usage has been deprecated; replace it with the usage above that includes the
<clear>
parameter.
:TRIGger:BLOCk:WAIT <blockNumber>, <event>, <logic>, <event>
:TRIGger:BLOCk:WAIT <blockNumber>, <event>, <logic>, <event>, <event>
The following table shows the constants for the events.
Trigger events
Event description
Event constant
No trigger event
NONE
Front-panel TRIGGER key press
DISPlay
Notify trigger block
<n>
(1 to 8); the trigger model generates a
trigger event when it executes the notify block
NOTify<n>
A command interface trigger:
•
Any remote interface: *
TRG
•
GPIB only: GET bus command
•
VXI-11: VXI-11 command
device_trigger
COMMand
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
<n>
(1 to 6)
DIGio<n>
Line edge detected on TSP-Link synchronization line
<n>
(1 to 3)
TSPLink<n>
Appropriate LXI trigger packet is received on LAN trigger object
<n>
(1 to 8)
LAN<n>
Trigger event blender
<n>
(1 or 2), which combines trigger events
BLENder<n>
Trigger timer
<n>
(1 to 4) expired
TIMer<n>
Source limit condition occurs
SLIMit
Example 1
:TRIGger:BLOCk:WAIT 9, DISP
Set trigger model block 9 to wait for a user to
press the TRIGGER key on the front panel
before continuing and to act on a recorded
TRIGGER key event that gets detected either
before or after reaching block 9.
Example 2
:TRIGger:BLOCk:WAIT 9, DISP, ENTer
Set trigger model block 9 to wait for a user to
press the TRIGGER key on the front panel
before continuing and to act only on a recorded
TRIGGER key event that gets detected when
block 9 is reached.
Also see
(on page 3-110)