Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-294
2461-901-01 A/November 2015
trigger.model.setblock() — trigger.BLOCK_NOP
This function creates a placeholder that performs no action in the trigger model; available only using remote
commands.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(
blockNumber
, trigger.BLOCK_NOP)
blockNumber
The sequence of the block in the trigger model
Details
If you remove a trigger model block, you can use this block as a placeholder for the block number so
that you do not need to renumber the other blocks.
Example
trigger.model.setblock(4, trigger.BLOCK_NOP)
Set block number 4 to be a no operation
block.
Also see
None
trigger.model.setblock() — trigger.BLOCK_NOTIFY
This function defines a trigger model block that generates a trigger event and immediately continues to the next
block.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(
blockNumber
, trigger.BLOCK_NOTIFY, trigger.EVENT_NOTIFY
N
)
blockNumber
The sequence of the block in the trigger model
N
The identification number of the notification; 1 to 8
Details
When trigger model execution reaches a notify block, the instrument generates a trigger event and
immediately continues to the next block.
Other commands can reference the event that the notify block generates. This assigns a stimulus
somewhere else in the system. For example, you can use the notify event as the stimulus of a
hardware trigger line, such as a digital I/O line.