Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-116
2461-901-01 A/November 2015
smu.digitize.limit[Y].low.value
This attribute specifies the lower limit for limit tests.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
−
1
Usage
lowLimit
= smu.digitize.limit[
Y
].low.value
smu.digitize.limit[
Y
].low.value =
lowLimit
lowLimit
The low limit value of limit
Y
(-9.11 to 9.11)
Y
Limit number:
1
or
2
Details
This command sets the lower limit for the limit Y test for the selected digitize function. When limit Y
testing is enabled, this causes a fail indication to occur when the measurement value is less than this
value.
Example
This example enables limits 1 and 2 for digitize voltage measurements. Limit 1 is checking for
readings to be between 3 and 5 V, while limit 2 is checking for the readings to be between 1 and 7 V.
The auto clear feature is disabled, so if any reading is outside these limits, the corresponding fail is 1.
Therefore, if any one of the fails is 1, analyze the reading buffer data to find out which reading failed
the limits.