Appendix D: Convert Model 2400 to Model 2461 code
Model 2461 Interactive SourceMeter® Instrument Reference Manual
D-10
2461-901-01 A/November 2015
Model 2400 command
:CALCulate2:NULL:ACQuire
Model 2461 command
[:SENSe[1]]:<function>:RELative:ACQuire
(on page 6-68), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
Notes
Model 2400 command
:CALCulate2:NULL:OFFSet
:CALCulate2:NULL:OFFSet?
Model 2461 command
[:SENSe[1]]:<function>:RELative
(on page 6-67), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
[:SENSe[1]]:<function>:RELative?
Notes
Model 2400 command
:CALCulate2:NULL:STATe
:CALCulate2:NULL:STATe?
Model 2461 command
[:SENSe[1]]:<function>:RELative:STATe
(on page 6-69), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
Notes
CALCulate3 subsystem
Model 2400 command
:CALCulate3:DATA?
Model 2461 command
Not available
Notes
Use reading buffers; see the
(on page 6-156).
Model 2400 command
:CALCulate3:FORMat
:CALCulate3:FORMat?
Model 2461 command
Not available
Notes
Use reading buffers; see the
(on page 6-156).
CONFigure
Model 2400 command
:CONFigure
:CONFigure?
Model 2461 command
Not available
Notes
DISPlay subsystem
Model 2400 command
:DISPlay:CNDisplay
Model 2461 command
(on page 6-42)
Notes
Model 2400 command
:DISPlay:DIGits
:DISPlay:DIGits?
Model 2461 command
(on page 6-40), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
:DISPlay:<function>:DIGits?
Notes
Settings are 3 to 6 instead of 4 to 7