![background image](http://html2.mh-extra.com/html/keithley/2461/2461_reference-manual_3993652834.webp)
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-161
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.math.format = smu.MATH_MXB
smu.measure.math.mxb.mfactor = 0.80
smu.measure.math.mxb.bfactor = 50
smu.measure.math.enable = smu.ON
Set the measurement function to voltage.
Set the math operation to mx+b.
Set the scale factor for the mx +b operation
to 0.80.
Set the offset factor to 50.
Enable the math function.
Also see
Calculations that you can apply to measurements
(on page 3-82)
(on page 8-158)