Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-158
2461-901-01 A/November 2015
smu.measure.limit[Y].low.value
This attribute specifies the lower limit for limit tests.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
−
1
Usage
lowLimit
= smu.measure.limit[
Y
].low.value
smu.measure.limit[
Y
].low.value =
lowLimit
lowLimit
The low limit value of limit
Y
(-9.11 to 9.11)
Y
Limit number:
1
or
2
Details
This command sets the lower limit for the limit
Y
test for the selected measure function. When limit
Y
testing is enabled, this causes a fail indication to occur when the measurement value is less than this
value.
Example
(on page 8-155).
Also see
(on page 8-154)
smu.measure.math.enable
This attribute enables or disables math operations on measurements for the selected measurement function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.OFF
Usage
value
= smu.measure.math.enable
smu.measure.math.enable =
value
value
The math enable setting:
•
Disable:
smu.OFF
•
Enable:
smu.ON
Details
When this command is set to on, the math operation specified by the math format command is
performed before completing a measurement.