
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2461-901-01 A/November 2015
3-47
Storing configuration index 4
Use one of the following methods to:
•
Set the instrument source level to 5.0 V
•
Store all active source settings to
MySourceList
as configuration index 4 by appending to the
end of the list
Using SCPI commands:
:SOURce:VOLTage:LEVel 5
:SOURce:CONF:LIST:STORe "MySourceList"
Using TSP commands:
smu.source.level = 5
smu.source.configlist.store("MySourceList")
The following figure shows the active state of the instrument after you change the source level to
Instrument settings stored in a source configuration list
(on page 3-35) for a complete
list of source settings that the instrument stores in a source configuration list.
Figure 109: Example configuration point 4
Recalling a configuration index
Use one of the following methods to recall configuration index 2 on
MySourceList
.
If recalling a configuration list separately, recall source configuration lists before measure
configuration lists. This order ensures that dependencies between source and measure settings will
be properly handled. You can also recall a source and measure list together using the recall
commands. See the recall commands in the command reference sections of this manual for details.