Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-106
2461-901-01 A/November 2015
Example
SOUR:CONF:LIST:CRE "VoltCustomSweepList"
SOUR:PULS:VOLT:LEV 1.17
SOUR:CONF:LIST:STOR "VoltCustomSweepList"
SOUR:PULS:VOLT:LEV 1.27
SOUR:CONF:LIST:STOR "VoltCustomSweepList"
SOUR:PULS:VOLT:LEV 1.33
SOUR:CONF:LIST:STOR "VoltCustomSweepList"
SOUR:PULS:VOLT:LEV 1.39
SOUR:CONF:LIST:STOR "VoltCustomSweepList"
SOUR:PULS:SWE:VOLT:LIST 0.03, OFF, "defbuffer1", 1, 3, 0, .005, ON,
"VoltCustomSweepList"
INIT
*WAI
TRAC:DATA? 1, 4, "defbuffer1", READ
Set the source function to voltage and create a source list called
CustomVoltSweep
.
Set the pulse level to 1.17 V and store that pulse level at the next index (1) in
CustomVoltSweep
.
Set the pulse level to 1.27 V and store that pulse level at the next index (2) in
CustomVoltSweep
.
Set the pulse level to 1.33 V and store that pulse level at the next index (3) in
CustomVoltSweep
.
Set the pulse level to 1.39 V and store that pulse level at the next index (4) in
CustomVoltSweep
.
Creates a pulse sweep using the
CustomVoltSweep
source list that steps through four pulse
levels three times, for a total of 12 pulses. Each pulse has a width of 30 ms and an off time of 5 ms.
Readings are not taken during this pulse sweep. Start the sweep at index 1. Then query the pulse
list.
Output:
1.17,1.00,1.00,1.00
Also see
(on page 3-64)
:SOURce[1]:<function>[:LEVel][:IMMediate][:AMPLitude]
(on page 6-93)
(on page 6-97)
(on page 3-107)