Section 3: Functions and features
Model 2461 Interactive SourceMeter® Instrument Reference Manual
3-64
2461-901-01 A/November 2015
Using TSP commands
reset()
smu.source.configlist.create("CurrListSweep")
smu.source.func = smu.FUNC_DC_CURRENT
smu.source.range = 100e-3
smu.source.level = 1e-3
smu.source.configlist.store("CurrListSweep")
smu.source.level = 10e-3
smu.source.configlist.store("CurrListSweep")
smu.source.level = 5e-3
smu.source.configlist.store("CurrListSweep")
smu.source.level = 7e-3
smu.source.configlist.store("CurrListSweep")
smu.source.level = 11e-3
smu.source.configlist.store("CurrListSweep")
smu.source.level = 9e-3
smu.source.configlist.store("CurrListSweep")
smu.source.sweeplist("CurrListSweep", 1, 0.001)
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.range = 20
trigger.model.initiate()
Increasing the speed of sweeps
To increase the speed of sweeps:
•
Reduce the NPLC.
•
Turn autozero off. If autozero is on, the instrument takes new reference and zero values for every
reading. This can slow down sweep operation. Be aware that if you disable autozero,
measurements may drift and become erroneous. To minimize drift when autozero is disabled, use
the autozero once feature. For more information on the autozero options, see
(on page 2-129).
Pulse operation
You may need to make pulsed I-V measurements to avoid self-heating effects, device damage, or to
observe the response of a device to a pulsed signal. Pulsed signals are created by turning the source
output on for a specified amount of time and then off for a specified amount of time.
Figure 115: Simplified pulse example