Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-97
:SOURce[1]:<function>:PROTection[:LEVel]:TRIPped?
This command indicates if the overvoltage source protection feature is active.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:SOURce[1]:VOLTage:PROTection[:LEVel]:TRIPped?
Details
When overvoltage protection is active, the instrument restricts the maximum voltage level that the
instrument can source.
If the voltage source does not exceed the set limits, the return is
0
. If the voltage source exceeds the
set limits, the return is
1
.
Example
SOUR:VOLT:PROT:TRIP?
If overvoltage protection is active, the output is:
1
Also see
(on page 2-121)
:SOURce[1]:<function>:PROTection[:LEVel]
(on page 6-96)
:SOURce[1]:<function>:RANGe
This command selects the range for the source for the selected source function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Source configuration list
Save settings
Source configuration list
Current: 1 µA
Voltage: 200 mV
Usage
:SOURce[1]:<function>:RANGe <n>
:SOURce[1]:<function>:RANGe?
:SOURce[1]:<function>:RANGe? DEFault
:SOURce[1]:<function>:RANGe? MINimum
:SOURce[1]:<function>:RANGe? MAXimum
<function>
The source function to which this setting applies:
•
Current:
CURRent
•
Voltage:
VOLTage
<n>
Set to the maximum expected voltage or current to be sourced; see
Details
for values; the ranges are:
•
Current: -10 A to 10 A
•
Voltage: -100 V to 100 V