Section 4: Source-measure considerations
Model 2461 Interactive SourceMeter® Instrument Reference Manual
4-30
2461-901-01 A/November 2015
Source reset values
Setting
Default value on
reset
Menu > Source > Config List > New List
:SOURce[1]:CONFiguration:LIST:CREate
(on page 6-83)
smu.source.configlist.create()
None (existing
configuration lists
deleted)
Menu > Source > Config List > System to Index
:SOURce[1]:CONFiguration:LIST:STORe
(on page 6-88)
(on page 8-182)
None (existing
configuration lists
deleted)
Not available from front panel
:SOURce[1]:<function>:DELay:AUTO
(on page 6-90)
(on page 8-176)
On
MENU > Source > Settings > High Capacitance
:SOURce[1]:<function>:HIGH:CAPacitance
(on page 6-92)
(on page 8-184)
Off
HOME > Source
:SOURce[1]:<function>[:LEVel][:IMMediate][:AMPLitude]
(on page 8-185)
0
HOME > Limit
:SOURce[1]:<function>:<x>LIMit[:LEVel]
(on page 6-94)
(on page 8-219)
Current: 105
µ
A
Voltage: 7.35 V
FUNCTION
key
(on page 6-95)
Not applicable for TSP
Current
MENU > Source > Settings > Overvoltage Protection
:SOURce[1]:<function>:PROTection[:LEVel]
(on page 6-96)
(on page 8-188)
None
MENU > Source > Settings > Source Range
(on page 6-97)
(on page 8-207)
Current: 1
µ
A
Voltage: 200 mV
HOME > Source Range
:SOURce[1]:<function>:RANGe:AUTO
(on page 6-98)
(on page 8-175)
On
MENU > Source > Settings > Source Readback
:SOURce[1]:<function>:READ:BACK
(on page 6-99)
(on page 8-208)
On
FUNCTION
key
(on page 6-95)
(on page 8-184)
Voltage
Buffer reset values
Setting
Default value on
reset
Not available from front panel
(on page 6-163)
(on page 8-22)
Default buffers:
Continuous
User-defined
buffers: Once
Not available from front panel
(on page 6-164)
(on page 8-25)
Default buffers: ON
User-created
buffers: OFF