Section 3: Functions and features
Model 2461 Interactive SourceMeter® Instrument Reference Manual
3-56
2461-901-01 A/November 2015
Setting up a sweep from the front panel
To set up a sweep from the front panel, you select options from the Sweep Settings screen.
Set up the sweep from the front panel
1. Select
FUNCTION
and select the source and measure functions.
2. On the Home screen, set the Source value.
3. Press the
Menu
key.
4. Under Source, select
Sweep
. The Sweep Settings screen is displayed.
Figure 113: Sweep Settings screen - first page
5. Make selections appropriate to your sweep. See the table below for detail on the options.
6. Swipe down to see additional options.
Figure 114: Sweep Settings screen - second page
7. Make selections appropriate to your sweep. See the table below for detail on the options.
8. Select
Generate
.
9. To run the sweep, press the
TRIGGER
key.