Event Master Devices
471
Note:
This cable connects the Genlock board to the Motherboard.
How to install the Genlock Assembly
To install the Genlock Assembly follow the same procedure in the reverse order.
15.19 Rear Rack Ears
Concerned parts
R767282K Rear Connector protector
Necessary tools
1 x Phillips Screwdriver #2
How to remove Rear Rack-Ears
1.
Remove the eight pan head screws (8-32x.31), four from each side, to detach the rear rack ears and
connector protectors from the unit.
Note:
You don
’
t need to remove the rear rack ears to access any other part of the unit.
Image 15-44
How to install Rear Rack-Ears
To install the Rear Rack-Ears follow the same procedure in the reverse order.
15.20 Front Rack-Ears
Flow chart
Image 15-45
Concerned parts
R767281K FRONT RACK EARS S3
Necessary tools
1 x Phillips Screwdriver #2
Summary of Contents for Event Master E2 Series
Page 1: ...ENABLING BRIGHT OUTCOMES User s Guide Event Master Devices...
Page 12: ...R5905948 12 Event Master Devices 12...
Page 24: ...R5905948 12 Event Master Devices 24 Safety...
Page 52: ...R5905948 12 Event Master Devices 52 General...
Page 82: ...R5905948 12 Event Master Devices 82 Hardware orientation...
Page 102: ...R5905948 12 Event Master Devices 102 Front Panel Menu orientation...
Page 272: ...R5905948 12 Event Master Devices 272 Image 6 137 EM GUI orientation...
Page 304: ...R5905948 12 Event Master Devices 304 Updating firmware...
Page 326: ...R5905948 12 Event Master Devices 326 Image 9 28 General operation example...
Page 382: ...R5905948 12 Event Master Devices 382 Preventive maintenance actions...
Page 444: ...R5905948 12 Event Master Devices 444 E2 Maintenance...
Page 528: ...R5905948 12 Event Master Devices 528 EC 200 EC 210 Maintenance...
Page 569: ...569 R5905948 12 Event Master Devices Troubleshooting C...
Page 572: ...R5905948 12 Event Master Devices 572 Troubleshooting...