GUI reference
R&S
®
ZND
392
User Manual 1173.9557.02 ─ 61
Query further cal unit properties:
SYSTem:COMMunicate:RDEVice:AKAL:DATE?
SYSTem:COMMunicate:RDEVice:AKAL:FRANge?
SYSTem:COMMunicate:RDEVice:AKAL:PORTs?
SYSTem:COMMunicate:RDEVice:AKAL:WARMup[:STATe]?
Temperature Compensation
Available for inline calibration systems (ICS) only. Applies to all inline calibration units
(ICUs) that are connected to the ICS and that are used with factory characterization
data.
If checked, the R&S
ZND firmware adjusts the factory characterization data according
to the ICU temperatures.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TEMPcomp
Calibration Type / Source
Selects the calibration type for the selected physical ports. For an overview, refer to
The reflection calibration types can be used for any combination of physical ports:
reflection calibrations are performed for each selected port.
A transmission calibration type requires at least two physical ports. For the
unidirec-
tional
transmission calibration types ("Trans Norm", "One Path Two Ports"), the direc-
tion (
"Source"
port) must be specified in addition.
Note:
For a unidirectional R&S
ZND, the bidirectional transmission calibration types
"Trans Norm Both", TOSM and UOSM are not available
Transmission normalization and "One Path Two Ports" calibration types require two-
port (Through) characterization data for the cal unit. For current calibration units, how-
ever, these two-port characterizations are typically not part of the factory characteriza-
tion. If a tooltip indicates missing two-port characterization data, simply perform a new
characterization of your cal unit. In the first dialog of the "Characterization" wizard,
select "Take All OSM and Through" to make sure that the necessary two-port data is
acquired. See also
Note:
For a unidirectional R&S
ZND, Through data can't be created using the charac-
terization wizard. Use a more powerful Rohde & Schwarz VNA to create two-port char-
acterizations.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
Common Port
This combo-box is only available if
"Calibration" tab of the "System Config" dialog.
It defines the port that must appear in each port assignment of an automatic calibra-
tion, and that will be used as the "center" of all Through measurements.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
Cal softtool
www.allice.de
Allice Messtechnik GmbH