GUI reference
R&S
®
ZND
471
User Manual 1173.9557.02 ─ 61
Figure 5-28: Fixture modeling dialog: EZD - Single Ended Ports
The fixture modeling proceeds in the following steps:
1. Measure one or more test coupons for the related fixture; see
The tools differ in the available coupon types.
2. Measure the DUT with the fixture; see
Chapter 5.13.5.3, "Measure DUT + Test Fix-
3. Run the selected tool to calculate the Touchstone files modeling the test fixture;
Chapter 5.13.5.4, "Save Fixture Model"
The calculated
sNp
file(s) can then be used to deembed the DUT at selected
port(s).
All files generated by actions in this dialog are stored in the directory C:\Users\Public
\Documents\Rohde-Schwarz\Vna\Embedding.
The current implementation assumes symmetrical lead-ins and lead-outs.
5.13.5.1
Common controls on the dialog
The following controls are located at the bottom of each dialog (ISD/SFD/EZD, single-
ended/balanced):
Offset Embed softtool
www.allice.de
Allice Messtechnik GmbH