GUI reference
R&S
®
ZND
366
User Manual 1173.9557.02 ─ 61
can be fulfilled. Changing the "Start Frequency" and "Stop Frequency" modifies the
"Freq Step Size".
●
If the sweep range is defined via "Center Frequency" and "Span Frequency", both
the "Span Frequency" value and the "Number of Points" can vary as the "Freq Step
Size" is changed. The "Span Frequency" is reduced as little as possible so that the
condition
"Freq Step Size" = "Span Frequency" / ("Number of Points" – 1)
can be fulfilled. Changing the "Span Frequency" modifies the "Freq Step Size".
Note:
●
This setting is valid for linear frequency sweeps only. It does not apply to logarith-
mic and segmented sweeps, power, time or CW mode sweeps.
●
Decreasing the "Freq Step Size" generally increases the measurement time.
Remote command:
Sweep Time / Auto
Varies the measurement time for a sweep or delays the start of each sweep.
●
"Sweep Time" is the total measurement time for the sweep. The minimum possible
sweep time is equal to the estimated value in "Auto" mode. Setting a larger sweep
time is equivalent to defining a
before each partial measurement.
●
"Auto" minimizes the sweep time. The
is set to 0 s. "Sweep Time" indi-
cates the estimated sweep time, depending on the current measurement settings.
The "Sweep Time" and "Meas Delay" values are maintained until changed explicitly
if "Auto" is switched off.
If a time sweep is active, "Sweep Time" cannot be modified. The analyzer uses the
previously defined sweep time settings.
Remote command:
[SENSe<Ch>:]SWEep:TIME
[SENSe<Ch>:]SWEep:TIME:AUTO
Meas Delay
Adds a delay time before the start of every partial measurement. See
"Partial measurements and driving mode"
"Meas Delay" is not available while a "Time" or "CW Mode" sweep is active.
As an alternative to increasing the delay (and thus the total sweep time), it is possible
to select "Alternated" driving mode; see
Remote command:
All Partial Meas'ments / First Partial Meas'ment
If
is set to a value > 0, this setting allows you to define how the measure-
ment delay is applied:
●
If "All Partial Meas'ments" is selected, the delay time is added before each partial
measurement. For a complete 2 port S-parameter measurement, the delay must be
added twice per sweep point.
●
If "First Partial Meas'ment" is selected, the delay time is added once per sweep
point only, irrespective of the measured quantities and the number of partial mea-
Sweep Softtool