Concepts and features
R&S
®
ZND
217
User Manual 1173.9557.02 ─ 61
4.7.16
Additional removable system drive
Option R&S
ZND-B19
Option R&S
ZND-B19 provides an additional removable system drive for the
R&S
ZND, including operating system and VNA firmware.
See the data sheet for the available variants of this option.
4.7.17
Extended power range
Option R&S
ZND-K7
For the R&S
ZND, software option R&S
ZND-K7 enables an extended power range of
-45 dBm to +3 dBm.
R&S
ZND: Recalibration Required
Although this is a software option for the R&S
ZND, the instrument has to be recalibra-
ted at Rohde & Schwarz service to comply with the specifications of the data sheet.
4.7.18
USB-to-IEC/IEEE adapter
R&S ZVAB-B44
Hardware option R&S ZVAB-B44 (order no. 1302.5544.03) comprises an adapter and
driver software for controlling external devices via IEEE 488 / IEC 625 (GPIB). The
driver software is installed on the network analyzer. Connect the USB port of the
adapter to any of the master USB connectors on the front or rear panel of the analyzer.
Connect the GPIB port of the adapter to the external device.
This hardware option is required, if you want to use GPIB for two purposes at the same
time:
●
to control the R&S
ZND from an external controller
●
to control external devices from the R&S
ZND
(via the USB2GPIB adapter)
To get the adapter to work, you have to install the NI VISA library on the R&S
ZND.