Command reference
R&S
®
ZND
780
User Manual 1173.9557.02 ─ 61
Parameters for setting and query:
<CircuitModel>
SL | SC
Possible circuit models (character data); see
Example:
*RST; :CALC:TRAN:VNET:GLO:DEEM:PAR:R? SC; R? SL
Query the default resistances for ground loop deembedding. The
response is 10000000; 0.
CALC:TRAN:VNET:GLO:DEEM:PAR:R SC, 2.2E+3
Increase the resistance for the Shunt C model to 2.2 kΩ.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding[:STATe]
<Boolean>
Enables or disables the deembedding function for ground loops. It is allowed to change
the circuit model and its parameters while deembedding is enabled.
Suffix:
<Ch>
.
Channel number.
Parameters:
<Boolean>
ON - Deembedding active
OFF - Deembedding inactive
*RST:
OFF
Example:
CALC:TRAN:VNET:GLO:DEEM:TND SL
Select the Shunt L circuit model for deembedding.
CALC:TRAN:VNET:GLO:DEEM:PAR:R SL, 2.2E+3; :
CALC:TRAN:VNET:GLO:DEEM ON
Increase the resistance for the Shunt L circuit model to 2.2 kΩ
and enable deembedding.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding:TNDefinition
<CircuitModel>
Selects the circuit model for ground loop deembedding.
Suffix:
<Ch>
.
Channel number.
Parameters:
<CircuitModel>
FIMPort | SL | SC | SG
Possible circuit models (character data); see
.
Example:
See
CALCulate<Ch>:TRANsform:VNETworks:GLOop:
Manual operation:
See
SCPI command reference
www.allice.de