Command reference
R&S
®
ZND
803
User Manual 1173.9557.02 ─ 61
Parameters:
<Boolean>
ON - embedding active
OFF - embedding inactive
*RST:
OFF
Example:
CALC:TRAN:VNET:SEND:EMB:TND CSL
Select the Serial C, shunt L circuit model for embedding.
CALC:TRAN:VNET:SEND:EMB:PAR:R2 CSL, 2.2E+3; :
CALC:TRAN:VNET:SEND:EMB ON
Increase the resistance R2 for the Serial C, shunt L circuit model
to 2.2 kΩ and enable embedding.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
TNDefinition
<CircuitModel>
Selects the circuit model for single ended port embedding.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<CircuitModel>
FIMPort | CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL |
SHLC | GSL | LSG | GSG | SLG | SGL | SGG
Possible circuit models (character data); see
*RST:
CSL
Example:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:WAVes
<Boolean>
Defines whether the wave-based, or the classical S parameter-based de-/embedding
calculation is used in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
ON (1)
Wave-based calculation (default)
OFF (0)
S parameter-based calculation
Manual operation:
See
SCPI command reference
www.allice.de
Allice Messtechnik GmbH