GUI reference
R&S
®
ZND
489
User Manual 1173.9557.02 ─ 61
When loading a touchstone file, the analyzer by default assumes odd ports left (VNA
side), even ports right (DUT side). However, it is also possible to instruct the analyzer
to assume an "Increasing Port Sequence", i.e. that the low port numbers are on the
VNA side. Furthermore it is possible to "Swap Gates", i.e. to instruct the analyzer to
swap the VNA and DUT side.
Note:
The loaded file is stored in the active recall set. Persisted recall sets contain the
full (de-)embedding data so that they can be transferred to other instruments.
Remote command:
5.13.8
Ground Loop tab
Allows you to specify a 1-port ground loop deembedding/embedding network.
Such a network is either defined
●
via its S-parameter stored in a one-port Touchstone file (
*.s1p
) or
●
by selecting a predefined lumped element model (Shunt L or Shunt C) and specify-
ing the available parameters (resistance/inductance or resistance/capacitance)
Use the complementary dock widget to activate or deactivate ground loop dembedding
and embedding (see
Chapter 5.13.1.6, "Ground Loop panel"
Background information
Refer to
Chapter 4.6.2.7, "Ground loop de-/embedding"
Offset Embed softtool