Command reference
R&S
®
ZND
864
User Manual 1173.9557.02 ─ 61
Example:
Alternative settings using the R&S ZVAB compatibility mode:
*RST; :SYSTem:LANGuage 'ZVABT'
*RST; :INIT:CONT OFF
Activate single sweep mode for all channels (including channel
no. 2 created later).
INIT:SCOP SING
State that a single sweep will be performed in the active chan-
nel.
CALC2:PAR:SDEF 'TRC2','S11'; :INIT2
Create channel no. 2 with a new trace and start a single sweep
in channel no. 2. Start a single sweep in the second channel.
Manual operation:
INITiate:CONTinuous:ALL
<Boolean>
Qualifies whether the analyzer measures in single sweep or in continuous sweep
mode.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Parameters:
<Boolean>
ON - the analyzer measures continuously in all channel, repeat-
ing the current sweep. The query returns ON (1) if at least one
channel is measured continuously.
OFF - the measurement is stopped after the number of sweeps
defined via
initiates a new measurement cycle.
*RST:
ON
Example:
Manual operation:
"All Channels Continuous/All Channels on Hold"
INITiate[:IMMediate]:ALL
Starts a new single sweep sequence in all channels. This command is available in sin-
gle sweep mode only (
OFF
). The data of the last sweep
(or of previous sweeps, see
Retrieving the results of previous sweeps
) can be read
CALCulate<Chn>:DATA:NSWeep:FIRSt?
SDATa, <count>
.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Note:
In contrast to all other commands of the analyzer, the
INITiate<Ch>[:IMMediate]...
commands have been implemented for overlap-
Chapter 6.4, "Command processing"
Example:
Usage:
Event
Manual operation:
SCPI command reference