GUI reference
R&S
®
ZND
359
User Manual 1173.9557.02 ─ 61
5.8.2
Power tab
The "Power" tab provides settings related to transmit and receive power. It is also dis-
played on the "Power Bw Avg" softtool.
5.8.2.1
Controls on the Power tab
Power
Determines the output power at the test ports for the sweep types "Lin Freq", "Log
Freq", "CW Mode", and "Time". Also determines the default output power for "Segmen-
ted" sweeps, where the output power can be set per segment.
The setting has no effect for
sweeps, where the source power is varied over a
continuous range.
Remote command:
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]
Start Power / Stop Power
Remote command:
SOURce<Ch>:POWer<PhyPt>:STARt
SOURce<Ch>:POWer<PhyPt>:STOP
RF Off All Channels
"RF Off All Channels" switches the internal power source for all channels off (if
checked) or on. Switching off the RF power helps to prevent overheating of a connec-
ted DUT while no measurement results are taken.
Remote command:
5.8.3
Time Domain X-Axis tab
If the active trace is a time domain trace, these settings define its stimulus axis.
Time domain analysis requires option R&S
ZND-K2. If this option is not installed, the
"Time Domain X-Axis" tab is hidden.
Stimulus softtool