Command reference
R&S
®
ZND
951
User Manual 1173.9557.02 ─ 61
[SENSe<Ch>:]CORRection:COLLect:AUTO:REPeat
Use this command to reuse measurement data of the active automatic calibration (if
available) for the next automatic calibration. This data reuse allows an automatic cali-
bration to be repeated without repeating all measurements.
Reusing existing calibration data is particularly useful for
, whose
OSM measurements can be repeated without reconnections, but whose unknown
through measurements can not.
Suffix:
<Ch>
.
Channel number
Example:
See
Chapter 8.2.5.5, "Inline calibration"
Usage:
Event
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
Calculates the system error correction for automatic calibrations with
, saves the data and applies the calibration to the active channel.
Requires successful calibration sweeps for all related port assignments (see
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
)
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:TEMPcomp
<Boolean>
Available if the calibration unit selected using
(ICS) only. Applies to all inline calibra-
tion units (ICUs) that are connected to the selected ICS and that are used with factory
characterization data .
If set to ON (1), the R&S
ZND firmware adjusts the factory characterization data
according to the ICU temperatures.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
Manual operation:
See
SCPI command reference