Command reference
R&S
®
ZND
792
User Manual 1173.9557.02 ─ 61
Parameters for setting and query:
<arg0>
STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC | LSSL |
SLCS | SCLS | SCCS | SLLS | GSSL | LSSG | SLGS | SGLS
Circuit model whose resistance R<i> shall be set, see
Example:
CALCulate<Ch>:TRANsform:VNETworks:PPAir:
Manual operation:
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>:PARameters:
G<1|2|3>
<arg0>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>:PARameters:
G<1|2|3>?
<arg0>
Specifies the conductance value G<i> in the different lumped circuit models for port
pair embedding.
In the query form, the <Conductance> parameter must be omitted. The command
returns the conductance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<ListId>
Index of the affected port pair (see
TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
DEFine
<1|2|3>
Index i of the conductance G<i> in the related lumped circuit
model. If unspecified the numeric suffix is set to 1.
Parameters:
<Conductance>
Conductance G<i> for the specified circuit model.
Range:
-1kS to 1 kS.
Increment: 1 pS (1E-12 S)
*RST:
0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<arg0>
STSG | SGST | GSSL | LSSG | GSSG | SLGS | SGLS | SGGS
Circuit model whose conductance G<i> shall be set, see
Manual operation:
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>[:STATe]
<Boolean>
Enables or disables the embedding function for
port set
<ListId>. It is allowed to
change the embedding network while embedding is enabled.
Suffix:
<Ch>
.
Channel number
SCPI command reference
www.allice.de
Allice Messtechnik GmbH