Command reference
R&S
®
ZND
749
User Manual 1173.9557.02 ─ 61
Options:
R&S
ZND-K3
Manual operation:
See
CALCulate:TRANsform:TIME...
The
CALCulate:TRANsform:TIME...
commands control the transformation into the
Chapter 4.7.5, "Time domain analysis"
CALCulate<Chn>:TRANsform:TIME:CENTer
.................................................................... 749
CALCulate<Chn>:TRANsform:TIME:DCHebyshev
............................................................ 750
CALCulate<Chn>:TRANsform:TIME:LPASs
......................................................................750
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam
.....................................................750
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:CONTinuous
................................. 751
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:EXTRapolate
................................ 752
CALCulate<Chn>:TRANsform:TIME:LPFRequency
........................................................... 752
CALCulate<Chn>:TRANsform:TIME:RESolution:EFACtor
.................................................. 752
CALCulate<Chn>:TRANsform:TIME:SPAN
.......................................................................752
CALCulate<Chn>:TRANsform:TIME:STARt
...................................................................... 753
CALCulate<Chn>:TRANsform:TIME:STATe
...................................................................... 753
CALCulate<Chn>:TRANsform:TIME:STIMulus
..................................................................754
CALCulate<Chn>:TRANsform:TIME:STOP
.......................................................................754
CALCulate<Chn>:TRANsform:TIME[:TYPE]
..................................................................... 755
CALCulate<Chn>:TRANsform:TIME:WINDow
...................................................................755
CALCulate<Chn>:TRANsform:TIME:XAXis
.......................................................................755
CALCulate<Chn>:TRANsform:TIME:CENTer
<CenterTime>
Defines the center time of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<CenterTime>
Center time of the diagram in time domain.
Range:
-99.99999999999 s to +99.99999999999 s
Increment: 0.1 ns
*RST:
1.5E-009 s
Default unit: s
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:CENT 0; SPAN 5ns
Set the center time to 0 ns and the time span to 5 ns.
Manual operation:
See
"Time Start / Time Stop / Time Center / Time Span"
Note:
If the x-axis is scaled in distance units (
CALCulate<Chn>:TRANsform:TIME:
DISTance
), then the center value is entered in m; the range and default value
changes accordingly.
SCPI command reference