Command reference
R&S
®
ZND
665
User Manual 1173.9557.02 ─ 61
Parameters:
<AttenLeadInScaling>
Manual operation:
See
CALCulate:FMODel:ISD:SCALe:FREQuency
<MaxFrequency>
Defines the maximum frequency to deembed.
Corresponds to the
max_frequency
batch mode parameter of the ISD tool.
Parameters:
<MaxFrequency>
Default unit: Hz
Manual operation:
See
CALCulate:FMODel:ISD:SCALe:FTIMe
<FltLeadInScalingTime>
Overrides the lead-in’s flight time in case the through-trace test coupon is a bit too
short or too long.
Sets/gets the
atten_scale 1
batch mode parameter of the ISD tool.
Parameters:
<FltLeadInScalingTime>
Manual operation:
See
CALCulate:FMODel:ISD:SMALlfixture
<Boolean>
Enables/disables the ISD tool's small fixture mode.
Parameters:
<Boolean>
Manual operation:
See
CALCulate:FMODel:ISD:TRACe:COUPling
<TraceCoupling>
Tells the ISD tool about coupling among lead-in traces.
Corresponds to the
coupling
batch mode parameter of the ISD tool.
Parameters:
<TraceCoupling>
NONE | WEAK | STRong
NONE: no coupling.
WEAK: coupling will be extracted even if there are 2 ports
enabled on the test coupon
STRong: if the test coupon is a 4-port file, and there are two
ports to be extracted, the ISD tool will optimize odd- and even-
mode insertion losses
Manual operation:
See
SCPI command reference
www.allice.de
Allice Messtechnik GmbH