Command reference
R&S
®
ZND
1017
User Manual 1173.9557.02 ─ 61
Manual operation:
See
7.3.14.14
[SENSe:]... (other)
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence
.................................................................... 1017
[SENSe<Ch>:]COUPle
<Mode>[, <NumberOfAlternatedBlocks>]
Determines the order of partial measurements and sweeps (see
tial measurements and driving mode"
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the sweep mode applies
to all channels in the active recall set.
Parameters:
<Mode>
ALL | AUTO | NONE | BLOCk
AUTO – automatic mode: fast sweeps are performed in alterna-
ted driving mode, slower sweeps in chopped mode
ALL – chopped driving mode, complete all partial measurements
before proceeding to the next sweep point
NONE – alternated driving mode on, reverse the order of partial
measurements and sweeps
BLOCk – block alternated driving mode
*RST:
NONE
<NumAltBlocks>
Defines the number of blocks for the "Block Alternated" driving
mode.
*RST:
3
Example:
COUP NONE
Activate the alternated driving mode.
TRIG:LINK 'PPO'
Set the triggered measurement sequence equal to one partial
measurement. Each trigger event starts one partial measure-
ment for all sweep points.
Manual operation:
See
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence
<RealPart>[, <ImaginaryPart>]
Specifies the complex reference impedance for the physical port numbered
<PhyPt>
(impedance renormalization).
Use
[SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe]
ured and default reference impedances.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
SCPI command reference
www.allice.de
Allice Messtechnik GmbH