Command reference
R&S
®
ZND
658
User Manual 1173.9557.02 ─ 61
RESonant
The 2x thru test coupon will be split and used directly for deem-
bedding (parameter value
3
).
This option may be more accurate when the fixture and 2x Thru
have the same impedance at every location
Manual operation:
"Test Coupons > Insertion Loss"
CALCulate:FMODel:ISD:COUPon:MEASure
For a fixture modeling with the ISD tool and coupon types SYMMetric2x | OPEN1x |
SHORt1x (see
CALCulate:FMODel:ISD:COUPon:TYPE
measurement of the coupon at the active ports (see
Example:
Chapter 8.2.6, "Fixture modeling"
Usage:
Event
Manual operation:
"Measure / Measure Open / Measure Short"
CALCulate:FMODel:ISD:COUPon:MEASure:FILename
<String>
For a fixture modeling with the ISD tool and coupon types SYMMetric2x | OPEN1x |
SHORt1x (see
CALCulate:FMODel:ISD:COUPon:TYPE
coupon properties from a Touchstone file.
Parameters:
<String>
Path to the Touchstone file, either absolute or relative to the
selected working directory (
).
Manual operation:
"Load File / 1x Open Preset / 1x Short Preset"
CALCulate:FMODel:ISD:COUPon:MEASure:FILename:CLEar
Invalidates the file path previously set using
CALCulate:FMODel:ISD:COUPon:MEASure:OPEN:
CALCulate:FMODel:ISD:COUPon:MEASure:SHORt:FILename
Usage:
Event
Manual operation:
"Load File / 1x Open Preset / 1x Short Preset"
CALCulate:FMODel:ISD:COUPon:MEASure:OPEN
For a fixture modeling with the ISD tool and coupon type OPSHort1x (see
CALCulate:FMODel:ISD:COUPon:TYPE
), this command starts the measurement of
the
Open
coupon at the active ports (see
Usage:
Event
SCPI command reference
www.allice.de
Allice Messtechnik GmbH