GUI reference
R&S
®
ZND
249
User Manual 1173.9557.02 ─ 61
The notation for wave quantities follows the usual scheme of the vector network ana-
lyzer:
●
The a-waves are the outgoing/transmitted waves at the analyzer's test ports.
●
The b-waves are the incoming/measured waves.
●
The source port for the stimulus signal must be specified in addition.
●
The port number range covers all test ports of the analyzer.
Wave Quantity
Selects the type (left pull-down list) and the port number assignment (right pull-down
list) of the
.
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "A1" ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "A1" ...
Source > Port
The source port for the stimulus signal: "Port 1" or "Port 2" for a bidirectional R&S
ZND.
For a unidirectional R&S
ZND: unidirectional vs. bidir-
on page 80) only port 1 may serve as source and hence this con-
trol is disabled.
The analyzer places no restriction on the combination of source ports and port num-
bers of the measured wave quantity. For example, it is possible to measure a
2
while
port 1 is the source port (e.g. to estimate the directivity of the internal test set's direc-
tional element).
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "A1" ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "A1" ...
Show as
Selects the physical unit of the displayed trace. It is possible to display the measured
"Voltage" V or to convert it into a power value P according to the formula
Meas softtool