GUI reference
R&S
®
ZND
543
User Manual 1173.9557.02 ─ 61
Show Cal Kit Label
Enables/disables the "Calibration Info" dialog during manual calibration (see
Independent of the state of the "Show Cal Kit Label" flag, cal kit labels are displayed in
several other manual calibration dialogs.
Use Reduced Number of Through
logic for every port assignment of an auto-
matic calibration.
Note that for this to work, the port assignments must have a star topology, with a
"Common Port" set using
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
Remote command:
[SENSe:]CORRection:COLLect:AUTO:MCONnect
Delete Cal Pool / Delete All Cal Kits
Deletes all calibration data and all cal kit data. See
Chapter 5.11.4.3, "Calibration Man-
Remote command:
n/a
Search Path for additional Cal Kits and Connector Types
Contains the name and path of a special directory for cal kit files (
*.calkit
). All cal
kit files in the special directory are loaded automatically as predefined kits (i.e. read-
only kits which cannot be modified) every time the VNA application is started. It is pos-
sible to select the default cal kit directory
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration\Kits
or
any other directory. "None" means that no additional cal kit files are loaded on start-up.
Use the special directory to make sure that you do not have to import kits manually,
even after terminating the VNA application improperly. In this case, previously imported
cal kit files are not stored in the recall set file.
Remote command:
User Interface tab
Provides general user interface configurations.
Setup softtool