
Command reference
R&S
®
ZND
947
User Manual 1173.9557.02 ─ 61
–
A VNA cal kit file name
*.calkit
with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
Note
that for
The suitable, ICU-specific characterizations must be specified
using
SYSTem:COMMunicate:RDEVice:AKAL:CKIT
Setting parameters:
<CalName>
Example:
SENSe1:CORRection:COLLect:AUTO:CONFigure
FNPort,''
Sets the automatic calibration to "Full n-port" with factory char-
acterization.
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
<PortNumber>
Defines the "Common Port" of an automatic calibration.
This port will be part of every port assignment that is defined using
]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
[SENSe:]CORRection:COLLect:AUTO:MCONnect
is
ON
, this port will be used as
the center of all Through measurements. If you set up the port assignments manually,
make sure that every port assignment contains it.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<PortNumber>
Number of the common port
Usage:
Setting only
Manual operation:
[SENSe:]CORRection:COLLect:AUTO:MCONnect
<MinimumConnect>
Enables or disables the
logic for every port assignment of an auto-
matic calibration.
Note that for this to work, the port assignments must have a star topology, with a
"Common Port" set using
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
Setting parameters:
<MinimumConnect> ON (1), or OFF (0)
Manual operation:
"Use Reduced Number of Through"
SCPI command reference
www.allice.de
Allice Messtechnik GmbH