Command reference
R&S
®
ZND
940
User Manual 1173.9557.02 ─ 61
Example:
Let's perform a full 3-port calibration with a two-port calibration
unit, using factory characterization and the port assignments
given in the table below.
Prerequisite: the adequate calibration unit was selected before
using
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
SENSe1:CORRection:COLLect:AUTO:ASSignment:
DELete:ALL
Deletes all available port assignments.
SENSe1:CORRection:COLLect:AUTO:CONFigure
'FNPort',''
Sets the automatic calibration to "Full n-port" with factory char-
acterization.
SENSe1:CORRection:COLLect:AUTO:ASSignment1:
DEFine 2,1,3,2
Creates port assignment 1 explicitely (no auto-detection).
SENSe1:CORRection:COLLect:AUTO:ASSignment2:
DEFine:TPORt 3,4
Creates port assignment 2 implicitely (auto-detection).
Before starting the calibration sweep for port assignment 1,
ensure test port 2 is connected to cal unit port 1 and test port 3
to cal unit port 2
SENSe1:CORRection:COLLect:AUTO:ASSignment1:
ACQUire
Perfoms the calibration sweep for port assignment 1
Before starting the calibration sweep for port assignment 2,
ensure test ports 3 and 4 are connected to the cal unit (in any
order)
SENSe1:CORRection:COLLect:AUTO:ASSignment2:
ACQUire
Perfoms the calibration sweep for port assignment 2; auto-
detects the existing port-connections at runtime
SENSe1:CORRection:COLLect:AUTO:SAVE
Checks whether the acquired calibration data are sufficient to
calculate the system error correction. If yes, applies them to the
selected channel.
Test Port
Port Assignment 1
Port Assignment 2
2
Cal Unit Port 1
-
3
Cal Unit Port 2
auto-detected
4
-
auto-detected
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
<TestPort1>[, <TestPort2>]...
Creates the default port assignment(s) for the specified test ports.
The number of created assignments can be queried using the
]CORRection:COLLect:AUTO:ASSignment:COUNt?
command.
SCPI command reference